Transport Ventilators

Transvent

Technical Specification

Parameter Specification
Equipment Classification Class III
Power Supply
A/C 100-240VAC, 50/60 Hz, 0.7-1.5A.
Battery Inbuilt Lithium Battery backup upto 5 hours.
Gas Supply Connection Hose DISS type quick disconnect plug male with maximum working pressure of 35 bar
Type Electronically controlled and pnuematically driven ventilator
Graphic Display
7 inch touch screen color TFT display; Resolution: 800 x 480 pixels
Waveform for Pressure vs. Time, EtCO2 vs Time
Dimensions 250 x 200 x 127 (LxWxH) in mm
Category Electronically controlled and pnuematically driven ventilator
Display Screen 7 inch touch screen color TFT display; Resolution: 800 x 480 pixels
Ventilation modes PPV, V-AC, V-SIMV, PCV, P-AC, P-SIMV, CPAP, Manual & CPR
Ventilation Monitoring Parameters Range Tolerance
Tidal Volume 50~2500 mL ±30mL or ±15% of the set value, whichever is greater
Respiratory Rate 0, 1~120 bpm ±1bpm or ±5% of the set value, whichever is greater
TI:E Ratio 9:1~1:9 ±10% of the set value
Pressure Control 5~60 cmH2O ±2cmH2O or ±10% of the set value, whichever is greater
Pressure Support 0, 3~35 cmH2O ±2cmH2O or ±10% of the set value, whichever is greater
PEEP 0~30 cmH2O ±2cmH2O or ±10% of the set value, whichever is greater
Pressure Limit 15~70 cmH2O ±2cmH2O or ±10% of the set value, whichever is greater
Activation Pressure (-20)~20 cmH2O ±1cmH2O or ±10% of the set value, whichever is greater
FiO2 40%, 100% ±10% of the set value
Pressure Rise Time Slow, Normal, Fast -
Ventilation Monitoring Parameters Range Tolerance
Tidal Volume 0~3000 mL ±30mL or ±15% of the set value, whichever is greater
Minute Ventilation Volume 0~80 L/min ±0.5L/min or ±15% of the set value, whichever is greater
Total Respiratory Rate 0~120 bpm ±1bpm or ±5% of the set value, whichever is greater
Peak Airway Pressure (-20)~100 cmH2O ±2cmH2O or ±10% of the set value, whichever is greater
Etco2 0~150 mmHg 0~40mmHg: ±2mmHg, 41~70mmHg: ±5%, 71~100mmHg: ±8%, 101~150mmHg: ±10%
Gas Supply Medical Oxygen 2.7 bar to 6 bar
Weight 3.4 Kg
Dimensions 260X200X127(LxHWxH) in mm
Environment Operating
Temperature: -20°C to 60°C
Humidity: 15% to 95%, non-condensing
Atmospheric Pressure: 70 kPa to 110 kPa